Verify the firmware version on the instrument to be the desired firmware version. When the update is completed, click the Disconnect button. During the firmware update process, the instrument will reboot.Ĩ. Browse to the correct firmware file that you have downloaded from step 1. ![]() To continue with the process, click the Accept and Install button.Ħ. In the desired module, click Update, and read the End-User License Agreement. Enter Instrument IP Address and Click Connect. Select GC and ALS, and then the required instrument model, in this example an 8890 GC. Notice: The Agilent GC Firmware Update Utility can be installed from the following resources: From A.com: GC Firmware Update, On the Agilent GC and GC/MS User Manual and Tools DVD set, which is shipped with all Agilent GC, Headspace, GC/MS instrument as part of an installation ship kit. Finally, it could also be found on the 8860/8890 GC Firmware Disk.ģ. ![]() (See Figure 1)įigure 1.Agilent GC Firmware Update in the Windows Menu Launch the Agilent GC Firmware Update Tool from Windows Start > Agilent Technologies folder. Notice: Please remember to take note of the folder location where you have saved and extracted your firmware file for Step 6 of this procedure.Ģ. Download the appropriate firmware files from GC Firmware Update page in. ![]() Do not interrupt communication, or turn off GC and computer during firmware update processġ. ![]() Caution: Always use complete downloaded firmware file.